Bench-mounted, primary standard, var. temp. source
LANDCAL P1600B2 is a new improved bench mounted, primary standard, variable temperature source designed for testing and precise calibration of most high temperature industrial radiation thermometers, thermographic equipment and thermocouples.
The LANDCAL P1600B2 is a new high stability, bench mounted primary temperature source designed for testing and precise alibration of most industrial infrared radiation thermometers at temperatures up to 1600 °C/2900 °F. The isothermal enclosure rovides a highly stable, uniform temperature along the length of the cavity, which is suitable for the calibration of thermographic nstruments and also the calibration of thermocouples by the comparison method.
The large diameter, conical ended, blackbody cavity is heated by six robust silicon carbide heating elements equally spaced round the cavity, providing continous, reliable operation over many years. The temperature of the cavity is measured using an ptional rare metal thermocouple supplied complete with traceable calibration certificate.
Temperatures | |||
---|---|---|---|
Low Temperature (C): | 500 | High Temperature (C): | 1550 |
Low Temperature (F): | 932 | High Temperature (F): | 2822 |
Technical Details
Maximum temperature: | 1600 °C/2900 °F |
Recommended temp.: | 500 to 1550 °C/950 to 2850 °F |
Heating rate: | 1.5 hours to 1400 °C/2550 °F |
Stability: | < ±0.5 °C/1.0 °F over 60 min at set temperature |
Radiation cavity: Dimensions: | Silicon carbide 50mm/2.0in dia. x 300mm/12in |
Sighting tube extension: | 49mm/1.9in dia. x 100mm/4.0in |
Emissivity: | 0.998 |
Heating elements (6): | SiC 151/356/20/25.4/7.4 |
Control thermocouple: | Pt 13% Rh/Pt, Type R |
Controller types Master: Over-temperature: | Eurotherm with RS232C serial interface Eurotherm |
Power requirement: | 220/240V a.c. 50 to 60Hz, or 415/240V or 380/220V (3 phase) |
Power consumption: | 7.0kVA, 2.3kVA per phase (3 phase) |
Measuring thermocouple: | Type B (6/30), R (0/13) or S (0/10) |
Dimensions: | 865 x 500 x 700mm/ 34.0 x 19.6 x 27.5in (L x W x H) |
Weight: | 62.0kg/136.6lb |
Suitable for the calibration of thermographic instruments and also the calibration of thermocouples by the comparison method.
Resource & Download
Calibration Furnaces – Overview | Download File (753 KB) |
Comments are closed.